ApneaLink

US and Latin America

22302

Canada

22303

1 Apnealink device

1 Program CD

3 Nasal Cannulas

2 AA Batteries

1 Reusable Belt

1 Carrying Case

1 Quick Software Setup Guide

1 USB download cable

ApneaLink Oximetry Accessories Kit

US and Latin America

22304

Canada

22308

1 Nonin XPOD oximeter

1 Sensor clip

3 Disposable Sensors

TECHNICAL SPECIFICATIONS
APNEALINK AND PULSE OXIMETER

Signal Recording

Breathing sounds
Respiratory flow
Blood oxygen saturation
Pulse
Battery voltage

Sampling Rates for the Channels

Respiratory flow/breathing sounds: 100 Hz
Saturation: 1 Hz
Pulse: 1 Hz
Battery: 1 Hz

Signal Processing

Signal recording: 20 Bit
Signal storage: 16 Bit

Internal Memory

Storage capacity: 15 MB
Recording period: 8 hours minimum

Power Supply to Recorder

2 NiMH rechargeable batteries: Mignon/AA/1.2V/
at least 2.1 Ah or
2 batteries: LR 6/Mignon/AA/1.5V/at least 2.1 Ah

Weight

Recorder (without rechargeable batteries or batteries):
Approximately 50 g (1.8 oz)
Pulse Oximeter: Approximately 30 g (1.1 oz)

Operating Conditions

Temperature: 20°C to 40°C (68°F to 104°F)
Humidity: 10% to 90% RH (non condensing)

Shipment/Storage Conditions*

Temperature: -20°C to +50°C (-4°F to +122°F)
Humidity: 10% to 90% RH
*Recorder without rechargeable batteries or batteries

Operating/Storage Air Pressure

800 hPa to 1060 hPa

Effective Range

Flow sensor: -10 hPa to +10 hPa
SpO

2

: +/- 2 digits

Pulse: +/- 3 digits

Interfaces

Nasal pressure cannula: Luer connection
Pulse oximeter: 3-pin Binder plug
Computer: Fullspeed USB 1.1

Dimensions

Recorder (length x width x height): 125 x 60 x 30 mm
(4.9" x 2.4" x 1.2")
Pulse oximeter: (length x width x height): 53 x 20 x 15 mm
(2.1" x 0.8" x 0.6")

Electromagnetic Compatibility

Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-
1-2, for residential, commercial and light industry environments. For further details, see “Guidance and
Manufacturer’s Declaration – Electromagnetic Emissions and Immunity” as follows.

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions and Immunity

The ApneaLink is intended for use in the electromagnetic environment specified below. The customer or user
of the ApneaLink should ensure that the system is used only in such an environment.

Emissions test

Compliance

Electromagnetic environment—guidance

RF emissions CISPR11

Group 1

The ApneaLink uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are unlikely to cause
any interference in nearby electronic equipment

RF emissions CISPR11

Harmonic Emissions
IEC 61000-3-2

Voltage fluctuations/
flicker emissions
IEC 61000-3-3

Class B

The ApneaLink is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage network that supplies buildings used for
domestic purposes.

Optional Accessories and Disposables
(US, Latin America and Canada)

Nasal Cannulas (25/pk)

70388

Nasal/Oxygen Cannula (25/pk)

70319

Belt, reusable

629052

Belt, Single Use (24/pk)

70406

Sensor, Oximeter-Single Use

70412

Sensor, Oximeter (Flex Wrap)-Reusable 1431002

Tape, Sensor

(for Flex Wrap Sensor), (25/pk)

70276

Sensor, Oximeter (Soft Sensor)-Reusable

70413

ORDERING INFORMATION & PRODUCT CODES

Medical Electrical Equipment needs special precautions with respect to EMC and needs to be installed and
put into service according to EMC information provided in this document.

Warnings: The ApneaLink should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the ApneaLink should be observed to verify normal, operation in the configuration
in which it will be used. The use of accessories other than those specified in this manual is not
recommended. They may result in increased emissions or decreased immunity of the ApneaLink.