Technical specifications, Storage – ResMed Nasal Mask Mirage Micro User Manual

Page 9

Advertising
background image

8

Technical Specifications

Note:

The mask system does not contain latex, PVC or DEHP materials.

Pressure–Flow

Curve

The mask contains passive venting to protect against

rebreathing. As a result of manufacturing variations, the vent

flow rate may vary.

Dead Space

Information

The dead space of the mask varies according to cushion size. It

is 128 mL for the Extra Large size.

Therapy

Pressure

4 to 20 cm H

2

O

Resistance

Drop in pressure measured (nominal)

at 50 L/min: 0.1 cm H

2

O

at 100 L/min: 0.7 cm H

2

O

Sound

DECLARED DUAL-NUMBER NOISE EMISSION VALUES in

accordance with ISO 4871. The A-weighted sound power level

of the mask is 30 dBA, with uncertainty 3 dBA. The A-weighted

sound pressure level of the mask at a distance of 1 m is

22 dBA, with uncertainty 3 dBA.

Environmental

Conditions

Operating temperature: +5°C to +40°C

Operating humidity: up to 95% relative humidity

non-condensing

Storage and transport: -20°C to +60°C

Storage and transport humidity: up to 95% relative humidity

non-condensing

Gross

dimensions

S:

140 mm (H) x 91 mm (W) x 93 mm (D)

M, L, LW, XL: 155 mm (H) x 91 mm (W) x 98 mm (D)

Mask fully assembled – no headgear.

Note:

The manufacturer reserves the right to change these specifications

without notice.

Storage

Ensure that the mask is thoroughly clean and dry before storing it for any
length of time. Store the mask in a dry place out of direct sunlight.

Pressure

(cm H

2

O)

Flow

(L/min)

4

19

8

28

12

34

16

40

20

45

0

5

10

15

20

25

30

35

40

45

50

4

6

8

10

12

14

16

18

20

Vent Flo

w R

ate (L/min)

Mask Pressure (cm H

2

O)

Advertising