Technical specifications – ResMed Positive AirwAy Pressure Device S8 Series II User Manual

Page 16

Advertising
background image

14

Technical Specifications

.

The manufacturer reserves the right to change these specifications without notice.

The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:

Settings Invalid Remove Card

The identification details on the
Data Card do not match the
identification details on your device.

Contact your clinician immediately.

Problem/Possible Cause

Solution

Operating pressure range

4 to 20 cm H

2

O

Dimensions (H x W x D)

112 mm x 164 mm x 145 mm

Weight

1.4 kg

Operating Temperature

+5°C to +36°C

Power Supply

Input range for S8 Series II with H4i:100–240V, 50/60Hz; 110V,
400Hz; 2.5A < 140VA (110W) (maximum power consumption).
Instantaneous peak power consumption <340VA.

Refer to the DC-12 converter instructions for DC ratings. Only
use the ResMed DC-12 converter for DC input.

Typical power consumption at 20 cm H

2

0 is 39VA or 20.7W.

Operating Humidity

10%–95% non-condensing

Storage and Transport
Temperature

-20ºC to +60ºC

Storage and Transport
Humidity

10%–95% non-condensing

Operating Altitude

Sea level to 2,591 m

Electromagnetic
Compatibility

Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments.
The electromagnetic compatibility tables for these ResMed devices
can be found on www.resmed.com, on the Products page under
Service & Support. Click on the PDF file for your device.

Air Filter

Two-layered, powder-bonded, polyester non-woven fiber

Air Tubing

Flexible plastic, 2 m

Air Outlet

The 22 mm conical outlet complies with EN 1281-1

IEC 60601-1 Classifications

Class II (double insulation), Type CF

Housing Construction

Flame retardant engineering thermoplastic

DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:

Sound Pressure Level

24 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1: 2002
26 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1: 2007

Sound Power Level

34 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1: 2007

Supplemental Oxygen

Recommended maximum supplemental oxygen flow: 4 L/min

Advertising
This manual is related to the following products: