Indication

Error

Cause

Rectification

Correct Cuff

pressure

reading

low

not
possible

Movement

Very weak
artery pulse

Cuff pressure
too high

Low

Batteries weak/ Fit four new identical 1.5V

voltage

exhausted

alkaline batteries type AA
LR6

Display empty

• Use a soft, slightly moistened cloth to

clean the OMRON MX3 Plus.
Caution! Do not use petrol, thinners or
similar solvents!

• Use a soft, moistened cloth and soap to

clean the cuff.
Caution! Do not wash the cuff!
Caution! Do not use petrol, thinners or
similar solvents!

• Do not carry out repairs of any kind

yourself. If a defect occurs consult your

OMRON retail outlet or distributor as
mentioned on the packaging.

Calibration
• It is recommended to have the OMRON

MX3 Plus inspected every two years to
ensure correct function and accuracy.
Consult your OMRON retail outlet or
OMRON distributor as mentioned on the
packaging.

Replace batteries
• Remove the battery out of the battery

6 Error: causes and rectifications

7 Maintenance and spare parts

Wait for at least three minutes to repeat
measurement.
Inflate the cuff 40 mmHg higher than
previous value.
Follow instructions, repeat
measurement.

Wait for at least three minutes to repeat
measurement.
Inflate the cuff 30 mmHg lower than
previous value.
Follow instructions, repeat
measurement.