• Wiring

A

B

C

D

– EXCITATION

+ EXCITATION

Gauge

Connector

Typical 4-20 mA Circuit

4-20 mA

CURRENT

LOOP

E

F

– EXCITATION

+ EXCITATION

Gauge

Connector

Typical 0-5 VDC Circuit

+ SUPPLY

VOLTAGE

+ OUTPUT

– OUTPUT

Typical 4-20 mA Circuit with Open Collector Alarms

R = 250 ohm for 1-5 V output

L

R = 500 ohm for 2-10 V output

L

DO NOT EXCEED 750 OHMS WITH 24 VDC SUPPLY.

NOTE: Alarm loads may be relay coils as shown below.

Connect 1 A 400 V diodes to protect from relay "kick".

Alarm 2

Alarm 1

To + Power

Gauge

Connector

Controller

Process

L

R

+Vin

A

B

C

D

E

F

A

B

C

D

E

F

4-20mA Loop

*Any voltage up

to 60 VDC @

160 mA max.

Alarm

Loads

(Optional but recommended

for surge protection)

+

Gnd

Alarm 1

Dig Gnd

Alarm 2

DC* Power

Supply

-

-Excitation

+Excitation

Process

Controller

L

R

+Vin

DPG10L

M3367/0606