Specifications – Omron Healthcare OMRON HEM-790IT User Manual

Page 58

Advertising
background image

58

Model:

HEM-790ITN (HEM-7080-ITZ2)

Display:

LCD Digital Display

Measurement Range:

Pressure: 0 to 299 mmHg, Pulse: 40-180/min.

Accuracy/Calibration:

Pressure: ±3mmHg or 2% of reading
Pulse:

±5% of reading

Inflation:

Automatic by electric pump

Deflation:

Automatic pressure release valve

Rapid Air Release:

Automatic exhaust valve

Pressure Detection:

Capacitive pressure sensor

Measurement Method: Oscillometric method
Pulse Wave Detection: Capacitance type pressure sensor
Power Source:

120V 60Hz AC adapter or 4 “AA” batteries

Battery Life:

Approx. 1400 uses when used 6 times a day
with 4 new alkaline batteries

Operating Temperature 50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
/Humidity:
Storage Temperature

-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH

/Humidity/Air Pressure: /700 - 1060 hPa
Main Unit Weight:

Approximately 14 7/8 oz. (420 g)
not including batteries

Main Unit Dimensions: Approx. 6 3/32" (l)

×

5 5/32" (w)

×

3 5/16" (h)

(155 mm

×

131 mm

×

84 mm)

Cuff Size:

Approximately 6" (w) (152 mm)
Cuff tube 23 5/8" (l) (600 mm)

Cuff Circumference:

Fits arm circumferences 9" to 17"
(220 mm to 420 mm)

Memory:

Up to 84 per user

Weekly Averages:

Up to 8 weeks

Contents:

Main Unit, ComFit

TM

Cuff, AC Adapter,

Instruction Manual, Quick Start Guide, 4 “AA” batteries,
Storage Case, USB Cable

NOTE: These specifications are subject to change without notice.

SPECIFICATIONS

HEM-790IT_EN_SP_r2.qxd 11/19/09 9:57 AM Page 58