WARNING:

This guide is only a summary of the steps necessary to properly disinfect your endoscope. Be sure to follow the

detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your endoscope when purchased.

To use this chart as a wallchart, cut
off the top of the chart as illustrated.

To insert this chart into a binder, punch out
the holes and fold the chart as illustrated.

Suction Cleaning Adapter

Tube to suction

To Suction Pump

Suction Cleaning

Adapter (MAJ-222)

To Suction Pump

Clean the bronchoscope

meticulously prior to disinfection.

Single-Use Suction Valve (MAJ-209)

Suction Cylinder

Single-Use Biopsy Valve (MAJ-210)

Remote Switches

Universal Cord

Instrument/Suction Channel Port

Electrical Connector

Water-Resistant Cap

(MH-553)

Venting Connector

Distal Tip

Bending Section

Light Guide

Boot

UP/DOWN Angulation Control Lever

Control Section

Insertion Tube

Alcohol

HIGH-LEVEL DISINFECTION

RINSING

TRANSPORTATION

DRYING

NOMENCLATURE

IMPORTANT

1

2

3

4

5

6

9

10

7

8

©

2

0

0

4

O

ly

m

pu

s

A

m

er

ic

a

In

c.

/

P

ri

nt

ed

in

U

SA

0

1

/0

4

P

W

C

-1

0

0

0

1

6

EXERA

BF-160 Series and EVIS

BF-240 Series Bronchovideoscopes

BRONCH

OSC

O

PE DISINFECTION GUIDE

Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information

WARNING:

This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the

detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.

Attach a syringe to the adapter.

Pull the plunger of the syringe to

fill the instrument/suction channels

and channel cleaning adapter

with disinfectant solution.

Fill basin with disinfectant. Attach a

suction cleaning adapter (MAJ-222)

to the previously cleaned broncho-

scope. Immerse completely in

disinfectant solution.

While immersed, detach the adapter

from the bronchoscope. Use a clean,

lint-free cloth to wipe away any bub-

bles clinging to external surfaces.

Soak all equipment according to

the time and temperature

recommended by the disinfectant

manufacturer. Use a timer to

measure the exact contact time.

Connect the suction cleaning

adapter and syringe. Inject air

to remove disinfectant from the

instrument/suction channels,

then remove the bronchoscope

from the disinfectant.

Disconnect the suction cleaning

adapter and immerse the

bronchoscope and adapter in

sterile water. Using a sterile,

lint-free cloth, thoroughly rinse

and wipe all surfaces.

Connect suction cleaning adapter

to bronchoscope and to a tube

from a suction pump. Aspirate

water through channels for 30

seconds. Remove from water and

aspirate air through channels for

60 seconds.

Turn the suction pump off. Hold the control

section with the openings pointing down and

disconnect the suction cleaning adapter.

Dry the bronchoscope and adapter with an

alcohol-moistened cloth. Make sure that the

equipment is completely dry prior to storage.

Hold all parts of the endoscope when carrying. Do

not let the ends of the bronchoscope swing freely.

• Wear appropriate personal protective

equipment during reprocessing.

• Autoclave all autoclavable accessories.
• Store the bronchoscope hanging vertically

with all valves and caps removed.

• Use a compatible disinfectant and test

it for proper concentration as recom-

mended by the manufacturer.

• Do not use the carrying case to store

"patient-ready" endoscopes. Any broncho-

scope placed in the carrying case must be

reprocessed prior to patient use.

Immerse the bronchoscope's distal

end in 70% alcohol and aspirate

alcohol for 5 seconds. Remove the

distal end from the alcohol and

aspirate air for 20 seconds.