2 Operation

2.1 Principle

The Omega FTB500 Series of Low Flowrate Meters has been
developed to meet the need for a low flow measurement device
for use with low to moderate viscosity clean liquids and for gas
measurement applications.

The Omega FTB500 Series of Low Flowrate Meters is a family of
low flow rate measurement devices whose design is based on a
Pelton Wheel-like rotor. The measured fluid is directed
tangentially through a velocity nozzle against the rotor causing it
to spin. The pickup coil senses the spinning motion of the rotor
through the housing and converts it into a pulsing electrical
signal. Summation of the pulsing electrical signal relates directly
to the total flow, while the frequency is related to the flow rate.

2.2 Precautions

Do not drop the meter. Dropping the meter may result in
damage to the meter housing and/or internals.

Do not operate the meter at flowrates greater than the
maximum flowrate marked on the meter. Operating at
flowrates greater than the maximum flowrate may over-
spin the meter. Over-spinning may result in damage to
the meter.

CAUTION

:

Avoid over-spinning the meter. Over-spinning the meter may
result in damage to the meter internals and lead to meter
failure.