CABLE TYPE F2

Red

=

+ Power

White

=

Common

Green

=

Output

CABLE TYPE C1

Red

= + Power

White = Common

Green = Output

HIRSCHMANN TYPE

PIN-1

=

+ Power

PIN-2

=

Common

PIN-4

=

Output

BENDIX 4-PIN, 6-PIN

Pin A

=

+ Power

Pin B

=

Output

Pin D

=

Common

Voltage Output Units 1-5, 1-6 Vdc

Current Output Units 4-20mA

Ratiometric (mV/V)

PX791 & PX792 SANITARY TRANSDUCERS –

ELECTRICAL CONNECTIONS

CABLE TYPE F2

Red

=

+ Power

White

=

– Power

Green

=

+ Output

Black

=

– Output

CABLE TYPE C1

Red

=

+ Power

White =

Power

Green =

+ Output

Black

=

– Output

HIRSCHMANN TYPE

Pin A

=

+ Power

Pin B

=

+ Output

Pin C

=

– Output

Pin D

=

– Power

BENDIX 4-PIN, 6-PIN

Pin A

=

+ Power

Pin B

=

– Power

Pin C

=

+ Output

Pin D

=

– Output

Pin E

=

Shunt Cal.

Pin F

=

Shunt Cal.

CABLE TYPE F2

Red

=

+ Power

Black

=

Power

CABLE TYPE C1

Red

=

+ Power

Black

=

Power

HIRSCHMANN TYPE

Pin-1

=

+ Power

Pin-2

=

Power

BENDIX 4-PIN, 6-PIN

Pin A

=

+ Power

Pin B

=

Power

WA R N I N G :

R E A D B E F O R E

I N S TA L L AT I O N

The diaphragm is very sensitive and fragile!
Do not let anything touch the diaphragm but
the fluid to be measured. Fluid hammer and
surges can destroy any pressure transducer
and must always be avoided. A pressure
snubber should be installed to eliminate the
damaging hammer effects. Fluid hammer
occurs when a liquid flow is suddenly
stopped, as with quick-closing solenoid
valves. Surges occur when flow is suddenly
begun, as when a pump is turned on at full
power or a valve is quickly opened.

Liquid surges are particularly damaging to

pressure transducers if the pipe is originally
empty. To avoid damaging surges, fluid lines
should remain full (if possible), pumps
should be brought up to power slowly, and
valves opened slowly. To avoid damage
from both fluid hammer and surges, a surge
chamber should be installed.

Symptoms of fluid hammer and surge's

damaging effects:

a) Pressure transducer exhibits an output

at zero pressure (large zero offset). If
zero offset is less than 10% FS, user
can usually re-zero transducer, install
proper snubber and continue
monitoring pressures.

b) Pressure transducer output remains

constant regardless of pressure.

c) In severe cases, there will be no

output.