3. Recommended Position

4. Installation

1

Preamplifier

2

Fitting
inlet

Hand

tighten

only!

In-line Applications

Preamplifier

1

2

Customer
supplied
pipe

3/4 in. NPT or
ISO 7/1-Rc 3/4

Submersible Applications

Sealant

PH-31542-TC Sensor Cap
(purchased separately)

A. O-ring lubricant
recommended
(factory applied)

A. O-ring lubricant
recommended
(factory applied)

Sealant

B. Align
marks

B. Align
marks

C.

C.

Fill with

3 to 4 in.

of sealant

3

• Vertical (0

°

) position optimum

• 4 fps or less for max. performance and sensor life

0

°

Process

pipe

+60

°

Customer supplied:
• 3/4 in. NPT threaded

pipe (A)

• J-box (B)
• Flex conduit (C)
• Quick release pipe

clamp (D)

• Tank flange (E)

A.

B.

C.

C.

E.

B.

A.

D.

Closed

tanks

Open

tanks

In-Line

Caution: If liquid level is not constant, always ensure

liquid contact with electrode tip

Customer supplied:
• 3/4 in. NPT nipple (male)
• J-box
• flex conduit

-60

°

3/4 in. NPT OR
ISO 7/1-Rc 3/4
female threads

2 in.

min.

Attach 3/4 in. watertight pipe to the top of the preamplifier. Secure
the threaded connection to prevent any leakage. For additional
defense against possible accumulation of condensation at the back
seal area of the sensor, fill the lower 3-4 inches (75-100 mm) of
conduit or extension pipe with a flexible sealant such as silicone.