Specifications – Omega Engineering ULTRASONIC LEVEL TRANSMITTER LVU-90 User Manual

Page 2

Advertising
background image

Step One

SPECIFICATIONS

Range:

3.6" to 6' (9 cm to 1.8 m)

Accuracy:

± .25% of span in air

Resolution:

0.125” (3 mm)

Frequency:

83 kHz (nominal)

Pulse rate:

2 pulses per second

Beam width:

8° conical

Dead band:

3.6” (9 cm)

Supply voltage:

12-36 VDC

Max loop resistance: 600 Ohms @ 36 VDC
Signal output:

4-20 mA, 12-36 VDC

Fail-safe diagnostics: Reverts to 22 mA

LED flashing

Temperature rating:

F: -40º to 140º
C: -40º to 60º

Temp. compensation: Automatic over entire range
Pressure rating:

30 psi (2 bar) @ 25 °C., derated @ 1.667 psi
(0.113 bar) per °C. above 25 °C.

Enclosure rating:

NEMA 4X / IP65

Enclosure material:

Polypropylene, U.L. 94VO

Transducer materials: PVDF
Mounting threads:

3/4” NPT for LVU-91
3/4” G for LVU-92

Mounting gasket:

Viton (3/4" G) LVU-92 only

Conduit connection: 1/2” NPT
CE compliance:

EN 50082-2 immunity
EN 55011 emission
(pending)

About Ultrasonic Technology:

An ultrasonic sound wave is pulsed five times per second from the
base of the transducer. The sound wave reflects against the process
medium below and returns to the transducer. The microprocessor
based electronics measures the time of flight between the sound gen-
eration and receipt, and translates this figure into the distance between
the transmitter and process medium below.

Radius

6’ Range

8

°

Cone

Temperature/Pressure Derating

35

30

25

20

15

10

05

00

-40 -20 00 20 40 60

Operating Pressure (psi)

Temperature (

°

C)

Unacceptable

Range

Acceptable

Range

Maximum Temperature/Voltage Derating

Continuous 20 mA Curve

100

80

60

40

20

0

Operating Voltage (VDC)

Ambient Sensor

T

emperature (

°

C)

Unacceptable

Range

Acceptable

Range

12 16 20 24 28 32 36

700

600

500

400

300

200

100

000

Max. Series Resistance (Ohms)

18 24 30 36

Supply Volyage (VDC)

Unacceptable

Range

Acceptable

Range

Electrical Loading Limits

POWER

4-20 mA

(-) (+)

GND

12-36 VDC

1

2

3

4

5

6

1 . 2

2 . 1

2 . 9

3 . 7

4 . 9

5 . 4

Radius

Inches

Range

Feet

3 . 1

5 . 2

7 . 3

9 . 5

11 . 6

1 3 . 7

Radius

cm

Advertising