Title, Title specifications – Omron MC 514 User Manual

Page 15

Advertising
background image

TITLE

15

TITLE

TITLE

TITLE

SPECIFICATIONS

Name:

Omron

®

IntelliTemp

Ear Thermometer

Model:

MC-514

Power source:

DC 3 Volt (lithium battery CR2032)

Battery life:

Approximately 2500 measurements

Sensor:

Thermopile

Temperature indication: In 4 digits, every 0.1°F or °C

Accuracy:

93.2 ~ 96.7°F

(34.0 ~ 35.9°C) :

±0.5°F

(±0.3°C)

96.8 ~ 102.2°F

(36.0 ~ 39.0°C) :

±0.4°F

(±0.2°C)

102.3 ~ 108.0°F

(39.1 ~ 42.2°C) :

±0.5°F

(±0.3°C)

Measurement range:

93.2 to 108°F

(34 to 42.2°C)

Weight:

Approx. 1.8 oz

(50 g)

(including the battery)

Operating

Ambient temperature: 50 to 104°F

(10 to 40°C)

temperature / humidity: Relative humidity: 30 to 85%

Storage

Ambient temperature: -4 to 140°F

(-20 to 60°C)

temperature / humidity: Relative humidity: 10 to 95%

Memory:

Memory function displays the previous reading

*ASTM Standards:

E 1965-98

Contents:

Thermometer, Storage Case, Instruction Manual,
and 21 Lens Covers

Accessories:

Omron MC-505LC Lens Covers

UPC:

0 73796 33514 4

The specifications may be changed without prior notice.

* ASTM laboratory accuracy requirements for the thermometer only (not including

a lens filter) in the display range of 96.8 ˚F to 102.2 ˚F (36 ˚C to 39 ˚C) for infared

thermomters is ± 0.4 ˚F (± 0.2 ˚C), whereas for mercury-in-glass and electronic

thermometers, the requirements per ASTM Standards E 667-86 and

E 1112-86 is ± 0.2 ˚F (± 0.1 ˚C).