Calibration procedure – Omega Speaker Systems HH303 User Manual

Page 14

Advertising
background image

Date JUN./06/2002

Calibration Procedure

Page

1 of 1

Model No.

HH302 & HH303

Environmental: 23 ± 5 ℃

K,J Thermometer (

HH302) calibrate procedure

1. Hold REL+AVG+TIMER then press power , release all button ,press HOLD +C/F in 5

seconds, LCD will show “CAL 1”.

2. Input 52.418 mV via copper wire , press REL button when LCD display a stable

value(about 14,000 count). Then LCD will display “CAL 2”

3. Input 33.290 mV via copper wire, press REL button when LCD display a stable

value(about 18,000 count), Then LCD will display “CAL 3”

4. Input 0 C via thermocouple wire, press REL button when LCD display a stable

value(less then 1000 count),

Now , press REL button again to finish calibrating .

K,J Thermometer (

HH303) calibrate procedure

1. Hold REL+AVG+ T1-T2 then press power , release all button ,press HOLD +C/F in 5

seconds, LCD will show “CAL 1”.

2. Input 52.418 mV via copper wire , press REL button when LCD display a stable

value(about 14,000 count). Then LCD will display “CAL 2”

3. Input 33.290 mV via copper wire, press REL button when LCD display a stable

value(about 18,000 count), Then LCD will display “CAL 3”

4. Input 0 C via thermocouple wire, press REL button when LCD display a stable

value(less then 1000 count),

Now , press REL button again to finish calibrating .

Omega

Advertising