How to measure blood pressure – Omron Healthcare HEM-907XL User Manual

Page 13

Advertising
background image

24

25

5. Place the stethoscope on the

patient's arm.

6. Push the START Button to start the

measurement.

• Do not push the START Button without

wrapping the cuff.

• Do not squeeze or press the cuff during

the measurement.

• If you want to inflate again after the start

of deflation, push the START Button.

• If you want to accelerate deflation after the

start of deflation, push the DEFLATION
(deflation control) / Measurement Results
Display Switch Button. Each time the
Button is pushed, cuff is deflated rapidly in
increments of 5 to 10 mmHg.

9. Push the ON/OFF (power) Button to

turn off the power.

7. Take the readings.

8. Push the STOP Button to remove air

inside the cuff.

• The unit does not automatically deflate in

the MANU Mode.

If the monitor determines that the pressure
value is not correct, an error display appears
(Er1 to 9). In this case, refer to Page 28 and
start the measurement again.

1. Push the ON/OFF (power) Button to

turn on the power.

3. Set the P-SET (inflation level setting)

Knob to "AUTO" or the target
pressure value.

4. Measure the patient’s arm size and

wrap appropriate cuff over the
patient's arm. (Refer to pages 14
and 15.)

2. Set the MODE Selector to "MANU".

HOW TO MEASURE

BLOOD PRESSURE

(IN MANUAL MODE)

HOW TO MEASURE

BLOOD PRESSURE

(IN MANUAL MODE)