Operation – Omega Speaker Systems OS641C-MV-LS User Manual

Page 9

Advertising
background image

7

OPERATION

1. Plug the infrared temperature transducer test leads into the Vdc input jack and

common or ground input jack on the DMM. Observe polarity.

2. Set the DMM to the 2V range only, ignore the decimal point.
3. Use " " button to select turn on or turn off the Laser Marker.
4. Point the lens at the object whose temperature is to be measured.
5. Press the MEAS button. Measurement is performed as long as the MEAS

button is kept pressed.

6. Referring to the spot size figure, aim the laser beam at the object whose

temperature is to be measured.

NOTE: Although the field of measurement (or Field of View) and the spot almost

coincide, actually the field of measurement corresponds to the diameter for
90% optical response. The object whose temperature is to be measured needs
to be larger than the measurement diameter (spot of size) by an adequate
margin at least 1.5 to 2 times larger.

7. Read the DMM diaplay.

Advertising