Omega PCL340 User Manual

Page 32

Advertising
background image

28

ST_START

Starts a switch test

ST_OPEN?

Returns pressure value at which the switch opened

ST_CLOSE?

Returns pressure value at which the switch closed

ST_DEAD?

Returns pressure value of the deadband of the switch

VAL?

Returns the measured values

ZERO_MEAS

Zeros the pressure module

ZERO_MEAS?

Returns the zero offset of the pressure module

Table 7: Parameter units

Units Meaning

MA

milliamps of current

V

Voltage in volts

PSI

Pressure in pounds per square-inch

INH2O4C

Pressure in inches of water at 4°C

INH2O20C

Pressure in inches of water at 20°C

CMH2O4C

Pressure in centimeters of water at 4°C

CMH2O20C

Pressure in centimeters of water at 20°C

FTH2O60F

Pressure in feet of water at 60°F

BAR

Pressure in bars

MBAR

Pressure in millibars

KPAL

Pressure in kiloPascals

INHG

Pressure in inches of mercury at 0°C

MMHG

Pressure in millimeters of mercury at 0°C

KG/CM2

Pressure in kilograms per square-centimeter

ON

Turn something on

OFF

Turn something off

UPPER

Upper Display

MIDDLE

Middle Display

LOWER

Lower Display

DCI

Current measure function

DCI_LOOP

Current measure function with loop power

DCV

Voltage measure function

P1

Port P1 pressure measure function

P1_ST

Switch Test with port P1 pressure measure function

Advertising