Toshiba Electromagnetic Flowmeter LF470/LF612 User Manual

Page 2

Advertising
background image

TIC-LF470B

2

Model LF470 Detectors

Fluid pressure: -15 to 150 psi, or -1.0 to 10 bar

(-0.1 to 1 MPa)

Principal materials:

Measurement tube – Alumina ceramics
Electrodes Platinum
Pipe connection port:

Standard316 stainless steel

Option Ti (titanium),

Polyvinyl chloride (shock-resistant)+
Ta (Tantalum) for grounding plate,
Polyvinyl chloride (shock-resistant)+
Pt-Ir grounding plate,
Hastelloy C (Equivalent)

Dedicated preformed cables:

Signal cable

– 2-wire shielded chloroprene

sheathed cable

Overall length: 0.28 inch (7mm)
Length: 0.20 inch (5m)

Excitation cable 3-wiren chloroprene sheathed

cable

Overall length: 0.28 inch (7mm)

Length:

0.20 inch (5m)

Coating: phthalic acid resin coating pearl-gray

colored

Weight: approximately 2.2 lb (1.0 kg)

(for each size excluding cables)

Model LF612 converters

Input signals

Analog signal the voltage signal from detector,

proportional to process flow rate (for LF612

separate type converter).

Digital input DI (opt.)

Signal type: 20 to 30Vdc voltage signal

Input resistance: 2.7k

Number of inputs: one point

DI function — One of the following functions can be

assigned to the optional DI signal.

Range switching — Selects either the higher or lower

range in the unidirectional or bidirectional 2-range

setting.

Totalizer control — Starts and stops the built-in

totalizer.

Fixed-value outputs —Outputs fixed-values for

current and pulse outputs.

Zero adjustment — Executes zero adjustment

(on-stream at zero flow rate).

Output signals

Current output:

4–20mAdc (load resistance 0 to 750

Ω)

Note: The current output cannot be used with the

PROFIBUS-PA communication.

Digital outputs — One point (std.) and one more

point is optionally available as follows.

Digital output DO1 (std.):

Output type: Transistor open collector
Number of outputs: One point
Output capacity: 30Vdc, 200mA maximum

Digital output DO2 (opt.):

Output type:

Solid-state relay output (non polarity)

Number of outputs: One point
Output capacity: 150Vdc, 150mA maximum

or 150Vac (peak to peak), 100mA maximum

DO1 and DO2 functions — One of the following

functions can be assigned to DO1 (std.) and/or
DO2 (opt.)

Pulse output (available only for DO1, DO2)

Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)

3.6 to 360,000 pulses/hr (DO2)

(Over 3,600,000 pulses/hr, auto-setting)

Pulse width: 0.5 to 500ms (but less than half of
the period for 100% flow rate)

Note: The same and simultaneous pulse is not

available between DO1 and DO2.)

Multi-range selection outputs (Note 1)
High, High high, Low, and/or Low low alarm

outputs (Note 2)

Empty pipe alarm output
• Digital Output Active Status (DO1 and DO2)

(Note 2)

• Preset count output
• Converter failure alarm output

Note 1: Two outputs (DO1 and DO2) are needed for

4-range switching and forward/reverse 2-range
switching.

Note 2:Normal Open (default set) or Normal Close is

selected for alarm outputs when programming.

When power failure occurs, unit will be fault to
Normal Open.