Operations 4

Operations 4

Operations 4

Operations 4

Operations 4

Teledyne Analytical Instruments

Operations

Operations

Operations

Operations

Operations

4.1

4.1

4.1

4.1

4.1

Filling the Reservoir

Filling the Reservoir

Filling the Reservoir

Filling the Reservoir

Filling the Reservoir

The reservoir is located on the right side of the analyzer case.

1) Insure that the cap on the drain spout is securely tightened.

2) Remove cap from fill port on top of reservoir.

3) Pour distilled water into reservoir until it is half full (about one

quart). The water will automatically flow into the humidifier
column.

4) Replace cap on fill port and securely tighten. A missing or loose

cap will permit the sample to vent into the analyzer case.

4.2

4.2

4.2

4.2

4.2

Detector Cell

Detector Cell

Detector Cell

Detector Cell

Detector Cell

The cell is located in the heated compartment on the left side of the

analyzer case, as shown in Figure 4-1. To open the compartment, unscrew
the captive knurled knobs at the top and bottom of the compartment and
remove the plastic window.

4.2.1

4.2.1

4.2.1

4.2.1

4.2.1

Cell Packaging

Cell Packaging

Cell Packaging

Cell Packaging

Cell Packaging

The cell is packaged separately from the analyzer. It is filled with

distilled water to prevent oxidation of the electrodes from exposure to the
atmosphere. The cell should be left filled with the distilled water until the
analyzer is installed and ready for operation. The cell should not be exposed
to the atmosphere for any prolonged duration.

4.2.2

4.2.2

4.2.2

4.2.2

4.2.2

Electrolyte

Electrolyte

Electrolyte

Electrolyte

Electrolyte

The cell electrolyte is Teledyne Type A, used in applications where

there is a complete absence of acidic anhydrides (CO

2

, SO

2

) in the sample

gas. Type A electrolyte is a 10% solution (w/v) of reagent-grade potassium
hydroxide (KOH) in distilled water.