TRENDnet MultiTemp III 18-1106-33 User Manual

Page 19

Advertising
background image

7. Detailed Description

Mains Switch. Starts the cooling compressor and the pump
motor that circulates the temperature controlled liquid.
Activates the heater if the temperature set is greater than the
actual temperature.

Change the temperature value. Values are changed in an
accelerating manner when either key is held down.

Changes between showing the actual temperature and

the various program modes.

Actual temperature mode; shows the actual temperature in
the bath.
S mode; sets the working temperature.
c mode; sets temperature correction value.
H mode; sets the maximum allowed set temperature.
8 mode; in 8 mode on, MultiTemp III must be manually re-
started using reset after power failure (See Section 7.5
Manual restart in mode 8).

Enters a set value if pressed before the display returns

to actual temperature value. The display changes from flas-
hing to constant to confirm the enter operation.

After the safety circuit has shut down MultiTemp III

and the error has been eliminated, press to re-start
the unit.

Warning! After a power interruption, the circulator restarts
automatically. Ensure that this cannot cause a dangerous situ-
ation.

Note: If you need to start MultiTemp III manually after

power failure by pressing reset, proceed directly to
Section 7.5 Operation mode 8.