Teledyne OXYGEN DETECTOR 319R User Manual

Page 17

Advertising
background image

13

319R

Appendix

TELEDYNE BROWN ENGINEERING

Analytical Instruments

Appendix

Specifications

Ranges:

0–200 PPM, 0–1999 PPM O

2

, and 0–

25% calibration range

Sensitivity:

1% of full scale

Signal Output:

Standard: 0–10 VDC, negative ground

4-20 mA, negative ground
plus contact closures for range
identification.

Display:

Digital 7 segment, 3½-digit LCD with
over range indicator.

Alarms:

One high alarm, adjustable & fail-safe.
One low alarm, adjustable & fail-safe.
One low alarm, fixed at 2% full scale and
fail-safe.

Accuracy:

±2% of full scale at constant temperature
and pressure of calibration
±5% of full scale throughout operating
temperature range (once temperature
equilibrium has been reached).

Response:

90% of full scale in 35 seconds.

Operating Temp. Range:

32 °F to 122 °F (0°C to 50°C)

Power:

85-264 VAC, 47-440 Hz or 10-36 VDC,
user specified, 10 W max.

Enclosure:

General purpose, panel mount

Cell Type:

Z-2

Cell Warranty:

6 months

Cell Life (Expected):

6 months

Dimensions:

2.75 "H × 6.0 "W × 2.75 "D
(63.5 mm × 152.4 mm × 63.5 mm)
Faceplate: 3.75 "H × 7.0 "W
(95.25 mm × 177.8 mm)
Probe unit: 4.0 "H × 6.0 "W × 3 "D
(101.6 mm × 152.4 mm × 76.2 mm)