Guidance and manufacturer’s declaration-electromagnetic immunity-

for all EQUIPMENT AND SYSTEMS

Guidance and manufacturer’s declaration-electromagnetic immunity

The ET1928L is intended for use in the electromagnetic environment specified below. The customer or the user

of the ET1928L should assure that it is used in such an environment.

Immunity test

IEC 60601

Compliance level

Electromagnetic environment-

test level

guidelines

Electrostatic

± 6 kV contact

± 6 kV contact

Floors should be wood, concrete or

discharge(ESD)

± 8 kV air

± 8 kV air

ceramic tile. If floors are covered with

IEC 61000-4-2

synthetic material, the relative humidity

should be at least 30%.

Electrical Fast

± 2 kV for power

± 2 kV for power

Mains power quality should be that of a

transient/burst

supply lines

supply lines

typical commerical or hospital

environment.

IEC 61000-4-4

± 1 kV for input/output

± 1 kV for input/output

lines

lines

Surge

± 1 kV line(s) to line(s)

± 1 kV line(s) to line(s) Mains power quality should be that of a

IEC 61000-4-5

± 2 kV line(s) to earth

± 2 kV line(s) to earth

typical commerical or hospital

environment.

Voltage dips, short

<5% U

T

<5% U

T

Mains power quality be that of a

interruption and

(>95% dip in U

T

)

(>95% dip in U

T

)

typical commerical or hospital

voltage variations

for 0.5 cycle

for 0.5 cycle

environment. If the user of the ET1928L

on power supply

requires continued operation during power

input lines

40% U

T

40% U

T

mains interruptions, it is recommended that

(60% dip in U

T

)

(60 % dip in U

T

)

the ET1928L be powered from an

IEC 61000-4-11

for 5 cycles

for 5 cycles

uninterruptible power supply or a battery.

70% U

T

70% U

T

(30% dip in U

T

)

(30% dip in U

T

)

for 25 cycles

for 25 cycles

<5% U

T

<5% U

T

(>95% dip in U

T

)

(>95% dip in U

T

)

for 25 cycles

for 25 cycles

Power frequency

3 A/m

3A/m

Power frequency magnetic fields

(50/60 Hz)

should be at levels characteristic of a

magnetic field

typical location in a typical commerical

or hospital environment.

IEC 61000-4-8

NOTE U

T

is the a.c. mains voltage prior to application of the test level.