Equipment required, Connector wiring, Performance verification procedure – Tektronix VX4353 User Manual

Page 61

Advertising
background image

VX4353

Page A – 16

Equipment Required

Table 1–1 lists the equipment required for the performance and verification procedure.

Table 1–1: Equipment Required

Required tools and equipment

Part number

VXI Mainframe (such as the Tektronix VX1410)

n/a

VXI Slot 0 with resource Manager (Tektronix VX4521)
and appropriate cables and interface cards.

n/a

Digital Multi Meter (DMM) with 4-wire Ohm capability, and
the ability to read more then 10 Gigohms.

standard lab equipment

Connector for VXI 1780S

n/a

Talker/Listerner (Send/Read)

n/a

Connector Wiring

1. Using the VX1780S connector, wire pins: 1, 3, 5, 7, 9, 11, 13, 15, 18, 20, 22, 24, 26,

27, 30, and 32 together, leaving a “pig tail” to connect to the DMM.

2. Wire pins: 2, 4, 6, 8, 10, 12, 14, 16, 19, 21, 23, 25, 27, 29, 31, and 33 together,

leaving a “pig tail” to connect to the DMM.

3. Plug this connector into P4 on the VX4353.

Performance Verification Procedure

4. Apply power to the VXI system and wait 10 seconds.

a. Only the green PWR LED will be lit.

b. Send “S

lf

” (Set all relays) to the VX4353.

5. Apply power to the VXI system and wait 10 seconds.

a. Only the green PWR LED will be lit.

b. Send “S

lf

” (Set all relays) to the VX4353.

6. Verify that all LED’s for CH00 through CH31 are lit.

7. Send “R

lf

” (Reset all relays) to the VX4353.

8. Verify that all LEDs for Ch00 through CH31 are out.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com