1 features – Teledyne MX300-I User Manual

Page 11

Advertising
background image

Portable Oxygen Monitor

Introduction

Teledyne

Analytical

Instruments

3

NOTE: IF THE LOW ALARM CONDITION IS TRIGGERED WITH

THE OXYGEN LEVEL STAYING ABOVE THE SAFETY LIMIT
OF 18%, THE ALARM BUZZER CAN BE DEACTIVATED BY
PRESSING THE ALARM SILENCE

()

KEY. IF ALARM

CONDITION IS NOT CORRECTED WITHIN ABOUT 115
SECONDS, THE ALARM BUZZER WILL REACTIVATE.
PRESSING THE ALARM SILENCE

()

KEY AGAIN WILL

SILENCE THE ALARM BUZZER ONCE AGAIN FOR ABOUT
115 SECONDS, WITH THE BUZZER REAACTIVATING IF
THE ALARM CONDITIONS PERSIST.

NOTE: IF THE LOW ALARM CONDITION IS TRIGGERED WITH

THE OXYGEN LEVEL FALLING BELOW 18%, THE

SENSOR SIGN WILL FLASH ALONG WITH THE GREEN

LIGHT, AND THE BUZZER WILL ACTIVATE. IN THIS CON-
DITION, THE ALARM BUZZER CANNOT BE SILENCED.

WHILE HOLDING DOWN THE ALARM SILENCE

()

KEY

MAY MOMENTARILY SILENCE THE ALARM BUZZER, AUD-
IBLE ALARM WILL RETURN UPON DISENGAGING THE
ALARM SILENCE

()

KEY. HOLDING THE KEY FOR

LONGER DURATIONS WILL INITIATE THE “STUCK KEY”
ALARM, WHICH CAN BE DEACTIVATED ONLY VIA REMOV-
ING THE BATTERIES FROM THE ANALYZER. ANALYZER
SHOULD BE RECALIBRATED UP ON REPOWERING.

1.1 Features

The following features are standard on the MX300-I Monitor:

Large easy to read 3½-digit LCD (see options)

Automatic LCD back lighting upon key press

Microprocessor controlled

Up/Down control keys on the front panel

Sensor fail/disconnect alarm indicator (audible and visual)

Two adjustable concentration alarms with user-defined set
points

Alarm Silence () key

2000 operating hours from 3 AA alkaline batteries

Battery status indicator

Battery test function

Stand for upright tabletop deployment