UEi UGI INF200 User Manual

Page 6

Advertising
background image

INF200-MAN

P. 5

S p e c i f i c a t i o n s

*Laser is offset 1 -2 cm (0.5”) above center of target spot.
**Battery life will vary based on backlight and laser storage.

Tro u b l e s h o o t i n g

No display or erratic display: check the battery for proper voltage and
tight contact at the battery clip. Ensure the unit is at the specified
operating temperature.

Constant or spurious over-load (OL) display: check battery voltage.
Check for electromagnetic interference (EMI). To check for EMI, move
the unit to an open area, away from high voltage and radio or radar
transmitting sources.

Errornous temperature readings: inspect the infrared lens for blockage
or contamination. Follow cleaning instructions. Check battery for proper
voltage and tight fit in the battery clip.

Emissivity Va l u e s

*Emissivity various with purity.

Temperature range

0 to 788˚F (-20 to 420˚C)

Accuracy

±2% of rdg. or ±3˚F (±2˚C), whichever is greater @
73 ±9˚F (23 ±5˚C) ambient operating temperature
and a known blackbody emissivity

Repeatability

±1% of rdg. or ±2˚F (±1˚C) whichever is greater

Response time

500 m sec. (95% response)

Special response

7-18 microns nominal thermopile detector

MIN/MAX/

T/AVG

X

temperature
Recall last reading

X

HI or LO

X

audible / visible alarm
LCD backlight

X

Laser (class 2 output)

X

Emissivity

.30 - 1.00 digitally adjustable

Temperature display

˚F or ˚C (selectable) 3 digit LCD

Display resolution

1˚F or 1˚C in all modes

Ambient operating range

32 to 120˚F (0 to 50˚C)

Relative humidity

10 - 95% RH noncodensing @ up to 85˚F (30˚C)

Storage temperature

-13 to 158˚F (-25 to 70˚C) without battery

Power

9V Alkaline battery

Battery life (Alkaline)

All models 50 hrs. (backlight not used)*
Enhanced laser sighting model 16 hrs.
Laser used 50%, backlight used 50%**

Dimensions

5.4” x 1.6” x 7.7” (137 mm x 41 mm x 196 mm)

Weight

9.5 oz. (270 gm)

Accessories

Water resistant hard case and manual

Distance spot ratio

8 to 1

Typical Emissivity Values - Metals
Surface

Emissivity

Iron and Steel
Cast iron (polished)

0.2

Cast iron (turned at 100˚C)

0.45

Cast iron (turned at 1000˚C)

0.6 to 0.7

Steel (ground sheet)

0.6

Mild steel

0.3 to 0.5

Steel plate (oxidized)

0.9

Iron plate (rusted)

0.7 to 0.85

Cast iron (rough) rusted

0.95

Rough ingot iron

0.9

Molten cast iron

0.3

Molten mild steel

0.3 to 0.4

Stainless steel (polished)

0.1

Stainless steel (various)

0.2 to 0.6

Aluminum
Polished aluminum

0.1*

Aluminum (heavily oxidized)

0.25

Aluminum oxide at 260˚C

0.6

Aluminum oxide at 800˚C

0.3

Aluminum alloys, various

0.1 to 0.25

Brass
Brass (polished)

0.1*

Brass (roughened surface)

0.2

Brass (oxidized)

0.6

Copper
Copper (polished)

0.1*

Copper plate (oxidized)

0.8

Molten copper

0.15

Lead
Lead (pure)

0.1*

Lead (oxidized at 25˚C)

0.3

Lead (oxidized, heated to 200˚C)

0.6

Nickel and its alloys
Nickel (pure)

0.1*

Nickel plate (oxidized)

0.4 to 0.5

Nichrome

0.7

Nichrome (oxidized)

0.95

Zinc (oxidized)

0.1*

Galvanized iron

0.3

Tin-plated steel

0.1*

Gold (polished)

0.1*

Silver (polished)

0.1*

Chromium (polished)

0.1*

Advertising