Welch Allyn PROPAQ CS User Manual
Page 142
Advertising
138
Fluke patient simulator modification
Welch Allyn Propaq CS Vital Signs Monitor
Incompatibilities with other simulators may also cause erroneous readings and/or missing
ECG. Check Propaq CS monitor readings with the simulator against a known source
before using the simulator for calibration.
Function
Color
Pin No.
Propaq CS 244/246
Output (+)
Black
4
C
Output (-)
Red
1
A & E
Exciter (+)
White
3
B
Exciter (-)
Green
5
D
ECG Ref
Blue
2
No Connection
810-0696-57
W1
W2