Microlife NEB 50 User Manual

Page 11

Advertising
background image

6

• The device is only allowed to be used for the purpose described in the instructions. The

manufacturer is not liable for damages due to incorrect usage.

• Do not subject the device to extreme temperatures, moisture, dust or direct sunlight.

• Never try to repair the device by yourself. Unauthorised opening of the device will

render the guarantee void.

• Ensure that children do not use the device without supervision; some parts are so

small that they could be swallowed and present a choking hazard.

• If the device has received a hard knock (fallen on the floor), you find a malfunction

or damage, or notice something unusual, you should have the device checked by

your national Microlife customer service organisation.

• Do not use in a ventilator circuit.

7. Guarantee

This device is

guaranteed for 2 years from the date of purchase. The guarantee

covers the device, not the replaceable components like the nebuliser, masks,

mouthpiece, air hose, and filter. The guarantee is void if the device is improperly used

or changes made by third parties. The guarantee is only valid on presentation of the

bill or a guarantee card completed by the dealer or pharmacy you purchased it from.

8. Technical specifications

Nebulisation rate:

0.3 ml/min. (3 ml inhalation takes less than 10 minutes)

Particle size:

1.9 µm (MMAD with 0.9% NaCl using API system) or

4.48 µm (MMD with Salbutamol using Malvern system)

Compressor air flow:

9 l/min.

Acoustic noise level:

58 dBA

Power source:

230 V~, 50 Hz

Power cord length:

2 m

Operating temperature: 10 - 40 °C; 10 - 95% relative humidity maximum

Storage temperature:

-25 - +70 °C; 10 - 95% relative humidity maximum

Operating limits:

30 min. ON / 30 min. OFF

Weight:

1560 g

Size:

195 mm x 160 mm x 105 mm

Reference to standards: EN60601-1, EN60601-1-2 (EMC)
Performance information provided above may vary for drugs supplied in suspension or

high viscosity form.

Advertising