Troubleshooting

Problems

Solutions

Compressor overheats. The compressor
has an internal thermal protector to
prevent the device from overheating. If the
internal temperature reaches the preset
level, the unit will automatically shut down,
and the ON/OFF button light will turn red.

Turn the unit OFF for a minimum of 40
minutes to allow the compressor to cool.

The compressor does not operate.

Check all power sources to the compressor.
Push the ON/OFF button again.

The device is noisy.

Check that the filter cover is correctly
positioned.

Medication does not nebulize properly.

Check for proper connection of the
tubing adapter.

Make sure the nebulizer is not empty.

Clean the nebulizer as described in
the manual.

Replace the nebulizer.

Check the intake filter and replace it if it
is dirty.

Compressor does not work, and the
ON/OFF button light is red.

The thermal protector has been activated
due to:

a. The device operating outside the
conditions outlined in this manual.

b. The device has been operating near
heat sources or in hot environments.

Allow the compressor to cool down for
a minimum of 40 minutes. Wait until the
red light turns off before turning the
compressor on again.

Specifications:

Power Supply:

100VAC - 120VAC~50/60 Hz

Weight:

0.9 lbs.

Size:

4.5" x 3.8" x 1.9"

Maximum pressure:

30 psi

Maximum flow rate:

8 LPM

Mode of operation

Intermittent use 30 min ON/30 min OFF

Class II Device (Double Safety Insulation).

Type B Device (device with specific protection against electrical hazards).

Certification:

Electric Safety Standards EN 60601-1

Electromagnetic Compatibility according to EN 60601-1-2

Storage Conditions:

Temperature:

MIN -10°F - MAX 158°F

Humidity:

MIN 10%RH - MAX 95%RH

Operating Conditions:

Temperature:

MIN 50°F - MAX 105°F

Humidity:

MIN 10%RH - MAX 95%RH

Replacement Parts and Accessories:

Filter 1036898

AC power/battery charger

1036899

12V DC (Car) Adapter [1.25 A (min.)] 1036519

Rechargeable Lithium Ion Battery

RDD489

Symbol Glossary:

Attention: Read accompanying documents

ON/OFF (power)

Alternating Current

Serial Number

~

SN

Optional Rechargeable Battery:

Operating time with battery

Up to 90 minutes

Number of treatments per charge

Up to 12 treatments when nebulizing 2.5 ml of medication

Normal time to charge

Approximately 6 hours