Activeforever Mettler Sonicator 740X Therapeutic Ultrasound Device User Manual

Sonicator, 740 specifications, General specifications

Advertising
background image

Sonicator

®

740 Specifications

General Specifications:

ETL and C-ETL Listed:

Model ME 740 (9801427)

Input:

100-240 VAC ±10%

Frequency:

60Hz

±3 and 50Hz ±3

Input Current:

less than 0.6 amps

Optional

Battery:

Rechargeable Smart Lithium Ion Battery Pack rated at 10.8Vand 4.8Ah

Patents:

U.S. and foreign patents applied for and granted including U.S. Patent

Numbers

4,966,131

and

5,095,890

Timer Accuracy:

±10% of the indicated time

Maximum treatment time:

29 minutes

Size:

13.5 in (L) x 9.5 in (W) x 7 in (H)

Weight:

3.75

pounds

without

battery

4.5

pounds

with

battery

Ultrasonic Generator Specifications:

Frequency:

1.0 MHz ±10%

3.2

MHz

±10%

3.3

MHz

±10%

Modes:

Continuous

Pulsed 10, 20 and 50% Duty cycle

Pulse repetition rate:

100 Hz ±5% (Pulse Mode)

Pulse duration:

1, 2 or 5 msec ±5%

Temporal peak/average
intensity ratio:

10:1, 5:1 and 2:1 ±5%

Maximum output power:

22 W with a 10 cm² applicator, (ME 7410)

11 W with 5 cm² applicator, (ME 7413)

2.2 W with a 1 cm² applicator (ME 7431)

Maximum intensity:

2.2 W/cm²

Indication accuracy:

±20% (for any level above 10% of maximum)

Ultrasonic Applicator Specifications:

Piezoelectric discs:

The output transducer utilizes a barium titanate disc with a specially

coated face.

Individual Applicator Specifications:

Applicator Part Number

Frequency

ERA

ME 7513

1 and 3.2 MHz ±10%

5 cm² ±20%

ME 7310

1 MHz ±10%

10 cm² ±20%

ME 7331

3.3 MHz ±10%

1 cm² ±20%

Maximum Beam
Uniformity

Ratio:

6:1

1333 South Claudina Street • Anaheim, California 92805 USA • Tel: 1 (714) 533–2221 • FAX: 1 (714) 635–7539

Advertising