Guidance and manufacturer's declaration – Activeforever Omron CompAir NE-C801 Nebulizer System (Omron CompAir Nebulizer System NE-C801) User Manual

Page 28

Advertising
background image

28

Electromagnetic immunity IEC60601-1-2

Immunity test

IEC 60601

Test level

Compliance

level

Electromagnetic environment –

guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

±6 kV
contact
±8 kV air

±6 kV
contact
±8 kV air

Floor should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.

Electrical fast
transient/burst
IEC 61000-4-4

±2 kV
for power
supply lines
±1 kV for
input/output
lines

±2 kV
for power
supply lines
*

1)

Mains power quality should be that of
a typical commercial and/or hospital
environment.

Surge
IEC 61000-4-5

±1 kV line
to line
±2 kV line
to earth

±1 kV line
to line
±2 kV line
to earth

Mains power quality should be that of
a typical commercial and/or hospital
environment.

Voltage dips, short
interruptions and
voltage variations on
power supply
IEC 61000-4-11

<5 % U

T

(>95 % dip
in U

T

) for

0.5 cycle

<5 % U

T

(>95 % dip
in U

T

) for

0.5 cycle

Mains power quality should be that of
a typical commercial and/or hospital
environment.
If the user of the NE-C801 requires
continued operation during power
mains interruption, it is recommended
that the NE-C801 be powered from
an uninterruptible power supply or
battery.

40 % U

T

(60 % dip
in U

T

) for 5

cycles

40 % U

T

(60 % dip
in U

T

) for 5

cycles

70 % U

T

(30 % dip in
U

T

) for 25

cycles

70 % U

T

(30 % dip in
U

T

) for 25

cycles

<5 % U

T

(95 % dip
in U

T

) for 5

sec.

<5 % U

T

(95 % dip
in U

T

) for 5

sec.

Power frequency
(50/ 60 Hz) magnetic
field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.

Note: U

T

is the A.C. mains voltage prior to application of the test level.

*

1)

The test of input/output lines is not applicable since they are shorter than 3.0m.

GUIdANCE ANd MANUfACTURER'S dECLARATION

NE-C801S-Z_A_M_E_SP_041311.indd 28

4/13/11 10:20:57 AM

Advertising