Question:

Can the 1860 or 1870N95 respirator be used in surgery?

Answer:

Yes. Both the 1860 and 1870 respirators are cleared for use as a surgical mask by the

FDA. In addition to being N95 respirators, they also provide greater than 99% bacterial

filtration efficiency against wearer generated particles according to the Modified Greene

and Vesley test method.They are also fluid resistant to help reduce exposure to blood

and body fluids. When worn properly and in combination with protective eyewear, they

comply with the OSHA Bloodborne Pathogens Standard. Fit testing is not required

when worn solely as a surgical mask.

Question:

Can the 1860 and 1870 respirators be used for laser and electrocautery procedures?

Answer:

Yes! In addition to recommended healthcare procedures that a smoke evacuator

should be used to reduce the amount of smoke in the environment, these respirators

may be used to further reduce the wearer’s exposure to electrocautery and laser plume,

offering higher filtration and increased facial seal over standard surgical masks.

Question:

Do respirators require fit testing?

Answer:

Yes! OSHA requires that a respiratory protection program be developed for the proper

use of these products as with all respirators when used for respirator protection. This

includes written procedures, medical screening, training and fit testing. 3M offers the

FT-10 Qualitative Fit Test Apparatus, which uses a saccharin aerosol, or the FT-30

Qualitative Fit Test Apparatus, which uses a bitter aerosol, for fit testing. These 10-20

minute fit test procedures demonstrate that an employee can attain a good face fit with

a given style respirator.