Activeforever Mettler Sys*Stim 294 4-Channel Neuromuscular Stimulator User Manual

Sys*stim 294 specifications

Advertising
background image

1

Sys*Stim 294 Specifications

General Specifications:

Input:

90–240 VAC, 50–60 Hz, 2.3 Amp. Nom.

ETL and C-ETL Listed:
Domestic Model

Model ME 294 (9801427)

Classification:

Protective Class I Equipment

CE model

Type BF Equipment

Enclosed equipment without protection against ingress of
water.

Equipment

not

suitable

for use in the presence of a

flammable anesthetic mixture with air or with nitrogen oxide

Year 2000 Compliant

Yes

Weight: 9.4

pounds

4,3 kg

Dimensions:

5 in (H) x 14.5 in (W) x 10 in (D)
12,7 cm (H) x 36,8 cm (W) x 25,4 cm (D)

Operating Temperature:

+50°F to +104°F
+10°C to +40°C

Humidity:

Operating, 30% to 75% Relative Humidity at 104

°F (40°C)

Nonoperating, 5 to 95% Relative Humidity, non-condensing

Storage Temperature:

-40°F to 167°F
-40°C to 75°C

Storage Humidity:

Storage, 30% to 90% Relative Humidity at 40º C, Non-
condensing

Storage Pressure:

700-900 mB

Environmental Disposal:

The device contains lead in the form of solder used to
produce electrical contact between components. To avoid
adverse environmental impact, utilize a disposal facility that
performs complete incineration of the device at a temperature
in excess of 1000°C.
The shipping materials are fabricated of cardboard and may
be disposed of with other paper products.

Treatment timer:

Timer Accuracy:

±0.5 minutes for times less than 5 minutes
±10% for times from 5 to 10 minutes
±1.0 minute for times greater that 10 minutes
±5%, CE specification

Maximum Treatment Time:

60 minutes–electrical stimulation

Advertising