Philips HeartStart OnSite Defibrillator

Philips HeartStart OnSite Defibrillator

Product specifications

Defibrillator

Defibrillator Model

HeartStart M5066A

Defibrillator Family

HS1

How Supplied

Defibrillator, Owner’s Manual, battery, 1 adult
SMART Pads cartridge, Quick Reference Guide and
Quick Start poster

Waveform

Truncated Exponential Biphasic.Waveform parameters
adjusted as a function of each patient’s impedance.

Energy

Single energy output. Adult: nominal 150 Joules into a
50 ohm load. Infant/Child: nominal 50 Joules into a 50
ohm load. Automatically set based on type of SMART
Pads cartridge installed.

Shock-to-Shock

Typically less than 20 seconds between shocks in a series.

Cycle Time

Quick Shock

Able to deliver a shock after the end of a CPR interval,
typically in eight seconds.

Voice Instructions

Detailed voice messages guide responder through use
of the defibrillator.

CPR Coaching

Instructions for adult and infant/child CPR available at
user’s option.

Shock Delivery

Via adhesive pads placed on patient’s bare skin as
illustrated on pads.

Controls

Green SMART Pads cartridge handle, green On/Off
button, blue i-button, orange Shock button

Indicators

Ready light; blue i-button; caution light

Physical Specifications

Size

2.8 x 7.4 x 8.3 inches (7 x 19 x 21 cm) H x D x W

Weight

With battery and pads case: 3.3 lbs. (1.5 kg)
Without battery or pads case: 2.4 lbs. (1 kg)

Environmental/Physical Requirements

Sealing

Solid objects per EN60529 class IP2X
Drip-proof per EN60529 class IPX1

Temperature

Operating: 32º - 122º F (0º- 50º C)
Standby: 50º - 109º F (10º- 43º C)

Humidity

Operating: 0% to 95% relative, non-condensing
Standby: 0% to 75% relative, non-condensing

Altitude

Operating: 0 to 15,000 feet
Standby: 0 to 8,500 feet > 48 hours and 8,500 to
15,000 feet < 48 hours

Shock/Drop Abuse

Withstands 1 meter drop to any edge,corner or surface.

Vibration

Meets EN1789 random and swept sine, road ambulance
specification in operating and standby states.

EMI (Radiated/Immunity)

Meets EN55011 Group 1 Level B Class B and
EN61000-4-3.

Patient Analysis System

Patient Analysis

Evaluates patient ECG to determine if a rhythm is
shockable. Rhythms considered shockable are
ventricular fibrillation (VF) and certain ventricular
tachycardias (VT) associated with lack of circulation.
For safety reasons, some VT rhythms associated with
circulation will not be interpreted as shockable, and
some very low-amplitude or low-frequency rhythms
will not be interpreted as shockable VF.

Sensitivity/Specificity

Meets AAMI DF80 guidelines and AHA
recommendations for adult defibrillation (Circulation
1997;95:1677-1682).

Artifact Detection

The effects of pacemaker artifact and electrical noise
are minimized with artifact detection.

Battery (M5070A)

Type

9 Volt DC, 4.2 Ah, composed of disposable long-life
lithium manganese dioxide primary cells.

Capacity

Minimum 200 shocks or 4 hours of operating time
(EN 60601-2-4:2003)

Install-by Date

Battery is labeled with an install-by date of at least five
years from date of manufacture.

Standby Life

Four years typical when battery is installed by the
install-by date. (Will power the AED in standby state
within the specified standby temperature range,
assuming one battery insertion test and no
defibrillation uses.)

SMART Pads

Adult SMART

M5071A defibrillation pads for patients 8 years of age

Pads Cartridge

and older or 55 lbs. (25 kg) and over.

Infant/Child SMART

M5072A defibrillation pads for patients under 8 years

Pads Cartridge

of age or 55 lbs. (25 kg). Rx only.

Energy Delivered

Adult: nominal 150 Joules into a 50 ohm load
Infant/Child: nominal 50 Joules into a 50 ohm load

How Supplied

Disposable cartridge, containing adhesive
defibrillation pads, clicks into defibrillator for an
integrated pads solution.

Active Surface Area

13.2 in

2

(85 cm

2

) each

Cable Length

Adult pads: 54 in (137.1 cm)
Infant/Child pads: 40 in (101.6 cm)

Use-by Date

Cartridge is labeled with a use-by date of at least two
years from date of manufacture.

Training Pads

Adult Training Pads

M5073A

Cartridge

Infant/Child Training

M5074A

Pads Cartridge

Function

Special pads put HeartStart OnSite into training mode
and disable its energy delivery capability.Training pads
feature 8 real-world training scripts. Used with training
mat (included) or with adapters on manikins.

Automated and User-activated Self-tests

Daily Automatic Self-tests

Tests internal circuitry, waveform delivery system, pads
cartridge and battery capacity.

Pads Integrity Test

Specifically tests readiness-for-use of pads
(gel moisture).

Battery Insertion Test

Upon battery insertion, extensive automatic self-tests
and user-interactive test check device readiness.

Status Indicator

Blinking green “Ready” light indicates ready for use.
Audible “chirp” indicates need for maintenance.

Data Recording and Transmission

Infrared

Wireless transmission of event data to a PC or Palm

®

PDA, using the IrDA protocol.

Data Stored

First 15 minutes of ECG and the entire incident’s events
and analysis decisions.

* Refer to the HeartStart OnSite Defibrillator Owner’s Manual for detailed product instructions.

All specifications based on 25º C unless otherwise noted.The defibrillator and its accessories are made

of latex-free materials.