PT2 MEASUREMENT MODES

Mode

Method of Measurement

Hold

once stabilized, the readings are captured then displayed

LIVE

real-time readings are displayed continuously during measurement

ACCESSORIES

STANDARD SOLUTIONS FOR CALIBRATION:

The ULTRAPEN™ PT2 requires pH 4, pH 7, and pH 10

buffer solutions for wet calibration and pH Sensor Storage

Solution for proper storage.

Order MODEL#s: PH4, PH7, PH10, and PHSS

REPLACEMENT PH SENSOR (with instructions):

Order Model: RPT2

MyRON L WARRANTy

The PT2, excluding the pH sensor, has a one (1) year

limited warranty.

The pH sensor has a six (6) month limited warranty.

Built On Trust. Founded in 1957, the Myron L Company is one of

the world’s leading manufacturers of water quality instruments.

Because of our commitment to product improvement,

changes in design and specifications are possible. You have

our assurance any changes will be guided by our product

philosophy: accuracy, reliability, and simplicity.

2450 Impala Drive

Carlsbad, CA 92010-7226 USA

Tel: +1-760-438-2021

Fax: +1-800-869-7668 / +1-760-931-9189

www.myronl.com

Printed in U.S.A.

© Myron L Company 2012

PT2DS 02-12

SPECIFICATIONS

pH Range:

0.00-14.00pH

pH Accuracy:

± 0.01pH (After Wet Calibration)

pH Repeatability:

± 0.01pH

pH Resolution:

0.01pH

Temperature Range:

0-71°C/32-160°F

Temperature Accuracy:

±0.1ºC/±0.1ºF

Temperature Repeatability:

±0.1ºC/±0.1ºF

Temperature Resolution:

0.1ºC/0.1ºF

Time to Reading Stabilization:

10-30 seconds

Power Consumption:

Active Mode 37mA, Sleep Mode 2µA

Temperature Compensation:

Automatic to 25ºC

Physical Dimensions:

17,15cm L x 1,59cm D or 6.75in. L x 0.625in. D

Weight:

54g/1.9oz. (without soaker cap and lanyard)

Case Material:

Anodized Aircraft Aluminum with Protective Coating

Battery Type:

N type, Alkaline, 1.5V

Calibration Solutions:

pH4, pH7, pH10

Operating/Storage Temperature:

0-55ºC or 32-131ºF

Enclosure Ratings:

IP67 and NEMA6

EN61236-1: 2006 – Annex A:

2008:

Electrostatic discharge to case of instrument may cause PT2 to spontaneously power on. If

this happens, the PT2 will power itself off within seconds.

Advertising