Veris Technologies OpticMapper - Field Reference Guide User Manual

Field reference guide —opticmapper

Advertising
background image

4. Install test load on OpticMapper external controller rear
panel

—port marked Optical Power port. Turn power on and go

to data acquisition mode on DataLogger or SoilViewer.

5. Clean sapphire window and install test reference under

window. Click OM System check on Soilveiwer, or press 2
on Datalogger for OM System Check. Follow onscreen
instructions. Reference values bellow.

Sensor 1: Difference of 200 or > between dark and light reference.
Sensor 2: Difference of 90 or > between dark and light reference.

6. Lower gauge wheels so EC

coulters and optical row unit
penetrate to desired depth.

7. Adjust top-link so implement

operates level in field position;
in rocky conditions shorten
top-link and tip implement
slightly forward to reduce
pressure on window; if soil
adheres to window, lengthen
top-link to increase pressure
on window.

OM values should be:
Sensor 1: 833
Sensor 2: 289

Readings are acceptable if
within 10 numbers from these.

Field Reference Guide

—OpticMapper

Keep with OpticMapper and refer to daily.

(Refer to Operating Instruction Manual for complete instructions)

1. Inspect unit for loose
or missing bolts, other
obvious problems

2. Do side depth wheels and
disks rotate freely? If not, clear
mud or other obstructions.

3. Check wear plate and window for excessive wear or
damage

STEP I. PRIOR TO MAPPING: PERFORM ROUTINE
DAILY MAINTENANCE

Field operations: (note: in-field adjustments may cause
shifts in sensor data)

Advertising