Veris Industries PD SERIES Install User Manual

Pd series, Notice, Installation guide

Advertising
background image

ZL0055-0D

PAGE 1

©2012 Veris Industries USA 800.354.8556 or +1.503.598.4564 / [email protected]

07121

Alta Labs, Enercept, Enspector, Hawkeye, Trustat, Veris, and the Veris ‘V’ logo are trademarks or registered trademarks of Veris Industries, L.L.C. in the USA and/or other countries.

TM

EnvironmEntal SEnSorS

inStallation GUiDE

Digital Pressure/Vacuum Gauge with Display

and Analog 4-20 mA Output

quick install

1. Select a location for the installation suitable to the gauge’s physical dimensions

that will allow access to the display and keypad. Plan for a ¼" M-NPT connection
or a hose to a ¼" fitting.

2. Thread the gauge onto the fitting. Apply tightening torque only to the metal

wrench flats, not to the PD housing. Use a 5/8” wrench. Do not overtighten.

3. Wire the device. See the wiring diagram in this guide.

4. Remove the protective film from the gauge front.

5. Program the gauge.

Installer’s Specifications

Measurement Units

Selectable: psi, bar, kg/cm2, ATM, in. Hg, in. H

2

O*

Accuracy**

< ±0.5% of F.S., BFLS

Overange Protection

2x rated pressure

Burst Pressure

5x rated pressure or 5000 psi, whichever is less

Temperature Ranges:

Operating (Ambient)

-10° to 70°C (15° to 158°F)

Storage

-40° to 65°C (-40° to 150°F)

Thermal Limits:

Compensated Range

0° to 55°C (32° to 131°F)

TC Zero

< ±1.5% of F.S.

TC Span

< ±1.5% of F.S.

Connection

¼" NPT, male

Power***

7.5 to 32 VDC

Housing

Polycarbonate

Environmental Protection

NEMA 4, IP65

Output

4-20 mA loop powered

* in. H2O units available on ≤250 psi range devices only

** Accuracy includes non-linearity, hysteresis and non-repeatability measured at 25°C/77°F

*** Select a loop power supply and total loop resistance such that when the loop current is 20 mA,

the gauge will have at least 7.5 VDC at its terminals.

PD SerieS

PD SerieS

Dimensions

NOTICE

• This product is not intended for life or safety applications.

• Do not install this product in hazardous or classified locations.
• Read and understand the instructions before installing
this product.
• Turn off all power supplying equipment before working on it.

• The installer is responsible for conformance to all applicable codes.

Range

= 17 - 4 SS

Output

= 4-20 mA

Material

50 = 0 to 50 psig
100 = 0 to 100 psig
250 = 0 to 250 psig
500 = 0 to 500 psig
1000 = 0 to 1000 psig

PD A

M

ProDuct iDentification

Zero

VERIS INDUSTRIES

http://www.veris.com

1-800-354-8556

1-503-598-4564

Digital Pressure Gauge

3.6”

(90mm)

4.2”

(106mm)

Calib

2.8”

(72mm)

1.8”

(46mm)

1.4”

(36mm)

2.2”

(56mm)

0.7”

(17mm)

Test Units

5/8”

Front

Side

Rear

Wrench Flat

(Tighten with a 5/8” wrench at this spot. Do not tighten by twisting the plastic PD housing.