Vegator 631 s1, Product description – VEGA VEGATOR 631 S1 User Manual
Page 3
VEGATOR 631 S1
3
Product description
VEGATOR 631 S1
The sensitivity of the meas. input was
increased to 2 MOhm. It is now also possible
to measure slightly conductive products.
Proceed as follows when connecting
VEGATOR 631 S1:
• Bridge terminals 3 and 4 on the plug-in
socket of VEGATOR 631 S1
• Open the housing of the conductive probe.
There must be no resistor between
connections 1 and 2. Should this be the
case, you have to solder it.
5
Bridge for special version S1
6
Terminal for probe
7
Function coding Ex-version
8
Instrument coding VEGATOR 631 Ex
9
Sockets for bridges
10 Transistor output
11 Relay output
12 Power supply
5
6
7
8
14
13
12
11
10
9
4
3
2
1
A
B
C
1
2
3
7
8
9
N
L1
-
+
5
6
7
8
10
11
9
12
Note:
The line monitoring and the fault signal are no
longer available. The measuring system must
be no longer used as partof an overfill
protection system acc. to WHG.