Adjustment with adjustment module minicom, Adjustment with the simatic pdm adjustment program, Product description – VEGA VEGAFLEX 54P (Profibus PA) User Manual

Page 8

Advertising
background image

8

VEGAFLEX 54P (Profibus PA)

Product description

2

4

4 ... 20 mA

-

+

ESC

OK

Tank 1
m (d)
12.345

-

+

ESC

OK

Tank 1
m (d)
12.345

Adjustment with detachable adjustment module. The
adjustment module can be plugged into the sensor or
into the external indicating instrument VEGADIS 50.

Detachable adjustment module MINICOM

-

+

ESC

OK

Tank 1

m (d)
12.345

max. 25 m

Adjustment with adjustment module
MINICOM

With the small (3.2 cm x 6.7 cm) 6-key ad-
justment module with display, you can carry
out the sensor-relevant adjustments directly
in the sensor.

The adjustment module can be plugged into
the sensor or into the optional, external indi-
cating instrument.

With the adjustment module MINICOM, you
can adjust the sensor and set the basic func-
tions. Beside setting the measurement condi-
tions and the simulation mode, it is also pos-
sible to assign the Profibus address and
carry out a false echo storage, see "4.2 Ad-
justment with the adjustment module
MINICOM“.

Adjustment with the SIMATIC PDM
adjustment program

To adjust all essential functions of the VEGA
sensor with the adjustment station SIMATIC
PDM from Siemens, a so-called EDD is re-
quired. Without this EDD, only the basic func-
tions such as min./max. wet adjustment or
integration time can be adjusted with the
PDM adjustment program. Further important
adjustment functions, such as the input of the
"

Meas. environment“ or a false echo storage,

are not available without EDD. After integra-
tion of the EDD files in the Simatic PDM ad-
justment software, all important adjustment
functions are accessible. If it is not at hand,
the obligatory GSD (instrument master file)
as well as the EDD (Electronic Device De-
scription) necessary for PDM can be
downloaded from the VEGA Homepage
(http://www.vega.com).

Advertising