Pressure transmitter D94 (Profibus PA)

29

Choose the requested measurement loop
and click to

OK

.

Setup

Pointer functions

Click in the window "Instrument data param-
eter adjustment" to

Meas. loop data

.

In the window "Meas. loop data" all available,
current sensor values and the peak values
(pointer function) are displayed. With the
button

reset

you can reset all displayed peak

values to the actual value all at once.

Simulation

For checking outputs of the pressure trans-
mitter and connected devices and compo-
nents, a pressure can be simulated. Click to

Diagnostics

, then to

Simulation

.

In the window "Simulation of outputs", you
click to

Start

to begin the simulation. With the

buttons "

<–

" and "

–>

" (or the thumb of the

scroll bar) you can adjust values between -
10 % and 110 %. Click to

Stop

to terminate

the simulation.

Note:

The simulation mode is not terminated auto-
matically, but remains active until you switch it
off!