3 adjustment – VEGA AM12 User Manual

Page 7

Advertising
background image

Indicating module 12

7

Set-up

3.3 Adjustment

• The indication required the adjustment of

the values for 4 mA and 20 mA. Proceed as
follows: Set the rotating switch to position
„POINT“ and push one of the keys. The
digital indication starts flashing. By pushing
the keys again you move the decimal point.
The minus key moves the decimal point to
the left, the plus key to the right.

• Set the rotating switch to position „ZERO“

and push the plus key to increase the
indication value or the minus key to reduce
the indication value. When you keep the
key pushed the digital indication modifies
the indication value in three steps with
increasing speed.

• Set the rotating switch to position „END“

and push one of the keys. The digital indi-
cation starts flashing. Use the keys to
adjust the indication value.

• Set the rotating switch to position „OPER-

ATE“. All adjusted values ar e transmitted to
an EEPROM-memory and saved. They
remain there even in case of a probable
voltage loss.

• Screw the housing cover .