4 reset – VEGA VEGADIS 371 Ex User Manual

Page 15

Advertising
background image

VEGADIS 371 Ex

15

relay module (see "4.2 Adjustment“).

4.4 Reset

You can carry out a reset on VEGADIS 371. Hence all
adjusted values are reset to the original factory setting.

• Interrupt the supply voltage of VEGADIS 371.
• Push the plus and minus key (3 and 4) together and

hold the two keys pushed. Switch on the supply
voltage. The display indicates "-RES-“. The
instrument carries out a reset.

VEGADIS 371 has the following fatory setting (Reset-
Default values).

Factory setting
Menu stage 1
- Relay switch on points

(10,0 %)

- Relay switch off points

(100,0 %)

- Display min.

(0)

- Display max.

(100)

- Decimal point

(1. position)

- Integration time

(0 s)

- Offset correction

(0 μA)

- Current output

(4 … 20 mA)

Menu stage 2
- Pump changeover function

(OFF)

- Working time indication

(0 hours)

Set-up

Advertising