4 dimensions – VEGA VEGADIS 50 Ex User Manual
Page 4
Advertising
4
VEGADIS 50
82
Pg 13,5
118
108
135
38
85
ø5
10
48
Product description
Digital indication
Zero point or final point
-9999 … 9999
Decimal point
individually selectable
Bargraph indication
0 % … 100 %
CE-conformity
VEGADIS 50 Ex meets the protective regulations of EMVG (89/336/EWG) and NSR
(73/23/EWG). The conformity has been judged acc. to the following standards:
EMVG
Emission
EN 50 081 - 1: 1993
Susceptibility
EN 50 082 - 2: 1995
NSR
EN 61 010 - 1: 1993
Deviating data for certified instruments
EC-type approval
TÜV 98 ATEX 1361 X
Classification
II 1G EEx ia IIC T6
Permissible ambient temperature
when used in Ex-atmospheres
Zone 0
Zone 1
- temperature class T6
55°C
55°C
- temperature class T5
60°C
70°C
- temperature class T4
60°C
85°C