4 dimensions – VEGA VEGADIS 50 Ex User Manual

Page 4

Advertising
background image

4

VEGADIS 50

82

Pg 13,5

118

108

135

38

85

ø5

10

48

Product description

Digital indication

Zero point or final point

-9999 … 9999

Decimal point

individually selectable

Bargraph indication

0 % … 100 %

CE-conformity

VEGADIS 50 Ex meets the protective regulations of EMVG (89/336/EWG) and NSR
(73/23/EWG). The conformity has been judged acc. to the following standards:
EMVG

Emission

EN 50 081 - 1: 1993

Susceptibility

EN 50 082 - 2: 1995

NSR

EN 61 010 - 1: 1993

Deviating data for certified instruments

EC-type approval

TÜV 98 ATEX 1361 X

Classification

II 1G EEx ia IIC T6

Permissible ambient temperature
when used in Ex-atmospheres

Zone 0

Zone 1

- temperature class T6

55°C

55°C

- temperature class T5

60°C

70°C

- temperature class T4

60°C

85°C

1.4 Dimensions

Advertising