9 supplement – VEGA VEGASWING 63 - Relay (DPDT) User Manual

Page 27

Advertising
background image

27

9 Supplement

VEGASWING 63 • - Relay (DPDT)

29229-EN-130409

Accuracy (according to DIN EN 60770-1)
Reference conditions and actuating variables according to DIN EN 61298-1

Ʋ Ambient temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Ʋ Product temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Product density

1 g/cm³ (0.036 lbs/in³) (water)

Ʋ Product viscosity

1 mPa s

Ʋ Superimposed pressure

0 kPa

Ʋ Sensor installation

Vertically from top

Ʋ Density selection switch

> 0.7 g/cm³

Measuring accuracy
Deviation

± 1 mm (0.04 in)

Influence of the process temperature on the switching point

1

3

4

2

10

(

25

/

64

")

8

(

5

/

16

")

6

(

15

/

64

")

4

(

5

/

32

")

2

(

5

/

64

")

-2

(

-5

/

64

")

-4

(

-5

/

32

")

-6

(

-15

/

64

")

-8

(

-5

/

16

")

-10

(

-25

/

64

")

0

0 °C

(32 °F)

100 °C

(212 °F)

150 °C

(302 °F)

200 °C

(392 °F)

250 °C

(482 °F)

50 °C

(122 °F)

Fig. 30: Influence of the process temperature on the switching point
1 Shifting of the switching point in mm (in)

2 Process temperature in °C (°F)

3 Switching point at reference conditions (notch)

4 Tuning fork