VEGA VEGACAL 62 Profibus PA User Manual

Page 40

Advertising
background image

40

10 Supplement

VEGACAL 62 • Profibus PA

30025-EN-130917

Input variable
Measured variable

level of non-conductive liquids and solids

Measuring principle

phase-selective admittance processing (PSA)

Measuring range

0 … 3000 pF

Measuring frequency

270 kHz

Accuracy (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1

Ʋ Temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Temperature error

Ʋ < 120 pF

< 1 pF

Ʋ > 120 pF

1 % of the current measured value

Linearity error

< 0.25 % of the complete measuring range

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
Process pressure

Ʋ Standard

-1 … +64 bar/-100 … 6400 kPa (-14.5 … 928 psig)

Ʋ with screening tube adapter (PN1)

0 … +1 bar/0 … 100 kPa (0 … 14.5 psig)

Process temperature VEGACAL 62 of

316L

-50 … +150 °C (-58 … +302 °F)

Process temperature (thread or flange

temperature) with temperature adapter

(option)

-50 … +200 °C (-58 … +392 °F)

1

2

3

-50°C

(-58°F)

50°C

(122°F)

40°C

(104°F)

-40°C

(-40°F)

80°C

(176°F)

0°C

(32°F)

100°C

(212°F)

150°C

(302°F)

200°C

(392°F)

Fig. 24: Ambient temperature - Process temperature
1 Process temperature

2 Ambient temperature

3 Temperature range with temperature adapter

Dielectric constant

≥ 1.5