6 Setup with a signal conditioning instrument

VEGACAL 63 • For connection to a signal conditioning instrument

30318-EN-130917

2. Carry out the adjustment on the signal conditioning instrument

(see operating instructions manual of the signal conditioning

instrument: "Adjustment with medium")

If the indication cannot be set to 100 %, proceed as follows:
If the indication does not reach 100 %, you have to set the range

selection switch on the probe one stage lower.
If the indication exceeds 100 % and cannot be reset, you have to set

the range selection switch of the probe to the next higher stage. In

both cases you have to repeat the adjustment.