VEGA VEGACAP 66 - two-wire User Manual

Page 29

Advertising
background image

29

9 Supplement

VEGACAP 66 • - two-wire

30023-EN-130916

Ʋ Single chamber housing

– 1 x IP 68 cable gland M20 x 1.5; 1 x blind plug

M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

Connection cable

Ʋ Wire cross-section

> 0.5 mm² (AWG 20)

Ʋ Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

Ʋ Tensile strength

< 1200 N (270 lbf)

Ʋ Standard length

5 m (16.4 ft)

Ʋ Max. length

1000 m (3280 ft)

Ʋ Min. bending radius

25 mm (0.984 in) with 25 °C (77 °F)

Ʋ Diameter approx.

8 mm (0.315 in)

Ʋ Colour - standard PE

Black

Ʋ Colour - standard PUR

Blue

Ʋ Colour - Ex-version

Blue

Adjustment elements
DIL switch for measuring range selection

Ʋ range 1

0 … 20 pF

Ʋ range 2

0 … 85 pF

Ʋ range 3

0 … 450 pF

Voltage supply
Operating voltage

10 … 36 V DC (via the signal conditioning instrument)

Electrical protective measures
Protection rating

Ʋ Plastic housing

IP 66/IP 67

Ʋ Aluminium and stainless steel stand-

ard

IP 66/IP 68 (0.2 bar)

2)

Ʋ Aluminium and stainless housing

(optionally available)

IP 66/IP 68 (1 bar)

Overvoltage category

III

Protection class

II

Functional safety (SIL)
Functional safety is already activated on instruments with SIL qualification ex factory. On instru-

ments without SIL qualification ex factory, the functional safety must be activated by the user via

the display and adjustment module or via PACTware for applications according to SIL.

Functional safety according to IEC 61508-4

Ʋ Single channel architecture (1oo1D) up to SIL2

2)

A suitable cable is the prerequisite for maintaining the protection rating.

Advertising