Breaking capacity
- min.

50 mW

- max.

750 VA AC, 54 W DC
If inductive loads or stronger currents are

switched through, the gold plating on the relay

contact surface will be permanently damaged.

The contact is then no longer suitable for

switching low-level signal circuits.

Contact material (relay contacts)

AgNi or AgSnO and Au plated

Modes (adjustable)

min./max.

Integration time
- when immersed

approx. 0.5 s

- when laid bare

approx. 1 s

Ambient conditions
Ambient temperature on the housing

-40 … +80 °C (-40 … +176 °F)

Storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions
Parameter

Limit level of solids

Process pressure

-1 … 6 bar/-100 … 600 kPa (-14.5 … 87 psi)

VEGAVIB 66 of 316L

-20 … +80 °C (-4 … +176 °F)

Density
- Standard

>0.02 g/cm³ (>0.0007 lbs/in³)

- adjustable

>0.008 g/cm³ (>0.0003 lbs/in³)

Electromechanical data
Cable entry/plug (dependent on the version)
- Single chamber housing

l

1x cable entry M20x1.5 (cable-ø 5 … 9 mm),

1x blind stopper M20x1.5, attached 1x

cable entry M20x1.5

or:

l

1x cable entry ½ NPT, 1x blind stopper

½ NPT, 1x cable entry ½ NPT

or:

l

1x plug M12x1, 1x blind stopper M20x1.5

Spring-loaded terminals

for wire cross-section up to 1.5 mm²

(0.0023 in²)