3 function chart – VEGA VEGAVIB 65 - contactless electronic switch User Manual

Page 20

Advertising
background image

With the potentiometer you can adapt the switching point to the

solid. It is already preset and must only be modified in special

cases.
By default, the potentiometer of VEGAVIB 65 is set to

complete right position (>0.02 g/cm³ or >0.0008 lbs/in³). In

case of very light-weight solids, you have to turn the

potentiometer to the complete left position (>0.008 g/cm³ or

>0.0003 lbs/in³). Hence the VEGAVIB 65 will be more

sensitive and can detect light-weight solids more reliably.
For instruments detecting solids in water, these values are not

applicable. The potentiometer is preset and must not be

changed.

With the mode adjustment (min./max.) you can change the

switching condition of the output. You can set the required

mode acc. to the "Function chart" (max. - max. detection or

overfill protection, min. - min. detection or dry run protection).
We recommend connecting acc. to the quiescent current

principle (contactless electronic switch is open when the

switching point is reached) because the contactless electronic

switch takes on the same (safe) condition if a failure is

detected.

Control lamp for indication of the switching condition.

l

green = switch closed

l

red = switch open

l

red (flashing) = failure

6.3 Function chart

The following chart provides an overview of the switching

conditions depending on the adjusted mode and level.

Level

Switching sta-

tus

Control lamp

Mode max.

Overfill protection

2

1

Switch closed

Green

Mode max.

Overfill protection

2

1

Switch open

red

Switching point adaptation (1)

Mode adjustment (2)

Signal lamp (5)

20

VEGAVIB 65 - with contactless electronic switch

Set up

29284

-EN
-060508

Advertising