VEGA VEGAVIB 61 - contactless electronic switch User Manual
Page 28
Process conditions
Measured variable
Limit level of solids
Process pressure
-1 … 16 bar/-100 … 1600 kPa (-14.5 … 232 psig)
6 bar
(87psi)
-50°C
(-58°F)
16 bar
(232psi)
50°C
(122°F)
100°C
(212°F)
0°C
(32°F)
150°C
(302°F)
200°C
(392°F)
250°C
(482°F)
1
2
Fig. 26: Process pressure - Process temperature
1
Process temperature
2
Process pressure
VEGAVIB
61 of 316L
-50 … +150 °C (-58 … +302 °F)
Process temperature (thread or flange tem-
perature) with temperature adapter (option)
-50 … +250 °C (-58 … +482 °F)
1
2
3
-50°C
(-58°F)
50°C
(122°F)
40°C
(104°F)
-40°C
(-40°F)
80°C
(176°F)
0°C
(32°F)
100°C
(212°F)
150°C
(302°F)
200°C
(392°F)
250°C
(482°F)
Fig. 27: Ambient temperature - Process temperature
1
Process temperature
2
Ambient temperature
3
Temperature range with temperature adapter
Density
-
Standard
>
0.05 g/cm³ (0.002 lbs/in³)
-
adjustable
>
0.02 g/cm³ (0.0007 lbs/in³)
Granular size
no limitation
1)
1)
max. 20 mm (0.8 in) with product density < 0.05 g/cm³ (0.002 lbs/in³).
28
VEGAVIB
61 • - contactless electronic switch
9 Supplement
29266
-EN
-120418