7 Maintenance and fault rectification

VEGACAL 65 • For connection to a signal conditioning instrument

30320-EN-130917

The electronics modules are adapted to the respective sensor. You

can find a suitable electronics module in the following overview.

CL-E60ZX (without approvals)

7.4 Shortening the electrode

The probe (cable) can be shortened by any length.

1. Loosen the two pins on the gravity weight (hexagon) and remove

the pins.

2. Pull the cable out of the gravity weight.
3. To avoid splicing of the steel cable, tin the cable before shortening

with a soldering iron and tighten the wire.

4. Shorten the cable with a cut-off wheel or metal saw at the lower

end. Make sure the length is correct before shortening.

70 mm (2.76"

)

200 mm

(7.87"

)

Fig. 14: Take the gravity weight into account and shorten the cable respectively

5. Shift the gravity weight again flush to the cable and fix it with the

two pins.

6. Carry out an adjustment. See "Setup procedure, carry out min.

adjustment - carry out max. adjustment".

7.5 How to proceed in case of repair

You can find a repair form as well as detailed information on how to

proceed under www.vega.com/downloads and "Forms and certifi-

cates".
By doing this you help us carry out the repair quickly and without hav-

ing to call back for needed information.
If a repair is necessary, please proceed as follows:

Print and fill out one form per instrument

Clean the instrument and pack it damage-proof

Attach the completed form and, if need be, also a safety data

sheet outside on the packaging

Assignment

Shortening the electrode