8 Maintenance and fault rectification

VEGACAL 62 • 4 … 20 mA/HART - two-wire

30024-EN-130917

Error

Cause

Rectification

E017

Adjustment span

too small

Carry out a fresh adjustment and in-

crease the distance between min. and

max. adjustment

E036

no operable sen-

sor software

Carry out a software update or send

instrument for repair

Depending on the reason for the fault and the measures taken, the

steps described in chapter "Set up" may have to be carried out again.

8.3 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.
In Ex applications only one instrument and one electronics module

with respective Ex approval may be used.
If there is no electronics module available on site, one can be ordered

from the VEGA agency serving you.

The new electronics module must be loaded with the order data of the

sensor. These are the options:

At the factory by VEGA

Or on site by the user

In both cases, the sensor serial number is needed. The serial num-

bers are stated on the type label of the instrument or on the delivery

note.
Information:

When loading on site, first of all the order data must be downloaded

from the Internet (see operating instructions manual "Electronics

module").

The electronics modules are adapted to the respective sensor and

differ in their signal output or in their power supply. You can find a suit-

able electronics module in the following overview.
The oscillators differ only in their signal output and are suitable for all

series 60 sensors.
The following types are available:

CL-E60H (4 … 20 mA/HART)

CL-E60P (Profibus PA)

CL-E60F (Foundation Fieldbus)

In Ex applications only an electronics module with respective Ex ap-

proval may be used.

8.4 Shortening the electrode

The probe (rod) can be shortened by any length.
1. Shorten the rod with a cut-off wheel or metal saw at the lower

end. Make sure the length is correct before shortening.

2. Carry out a fresh adjustment